0 Troubleshooting of Pipettes

The fourth step is to check the piston system. After disassembling the piston, it is required to check if the piston is polluted, clean the piston, smear lubrication oil, re-install the piston, and recheck if the pipetting volume is still low. If the problem can not be solved, please deliver to the supplier. If there are serious problems such as aging of the sealing ring or rust of the piston, please also send it to the supplier for handling. Generally, users are not recommended to replace the accessories after purchasing the accessories from the supplier.

Check whether the sealing device and piston are polluted. The piston should be glossy and free of corrosive rust. If it is contaminated, clean it with distilled water or isopropyl alcohol, and then wipe it with a non-woven cloth. At the same time, note that before re-installing, check whether the inner wall of the handle is clean and ensure that all parts are completely dry.

0 Function of Pipette in Laboratory

The pipette is a single-channel micropipette that is commonly used in biological and chemical laboratories to move liquids with a small volume, and it is a measuring tool for transferring liquid from the original container to another container within a certain range. The design of the pipette fully conforms to the ergonomic principle, and is characterized by easy disassemble and maintenance. The pipette has five modes: Pipetting mode, sample mixing mode, reverse aspiration mode, electrophoresis sampling mode, and graded pipetting mode. The pipette can be applied to liquids such as high viscosity liquids, high vapor pressure liquids, and foaming liquids. Pipettes are widely used in clinical diagnostic laboratories, biotechnology laboratories, pharmaceutical and chemical laboratories, environmental laboratories, and food laboratories.

0 How to Use the Pipette?

Before pipetting, it is required to make sure that the pipette, pipette tip and liquid are at the same temperature. In sucking liquid, the pipette must be kept in an upright position and the pipette tip shall be inserted 2-3 mm below the liquid surface. Before sucking the liquid, it is recommended to suck and release the liquid for several times to wet the suction nozzle (especially when it is going to suck liquid with a viscosity or when the density is different from water). At this time, two pipetting methods can be used:

The first is the forward pipetting method. Press the button to the first stop with the thumb and then slowly release the button to the original point(suck the liquid of the fixed volume), and then press the button to the first stop point to drain out the liquid, wait for a while, continue to press the button to the second stop point to blow out the residual liquid, and release the button in the end.